GUI

Calibration protocol

  1. input low_rpm

  2. input high_rpm

  3. input set_time (time of pump operation for a single set rpm)

  4. press "CALIBRATE"

  5. after set_time minutes:

    1. pop-up "low_rpm calibration complete"

    2. enter volume

    3. press continue

  6. after set_time minutes:

    1. pop-up "high rpm calibration complete"

    2. enter volume

    3. press done

    4. pop-up mL/RPM value

  7. calibration parameter (mL/RPM) stored to device (calib.json)

WARNING: calibration file not present, run calibration!

Single speed

Speed profile

The user can load a speed profile file.